Blog
Why Surgical Instruments Rust and Stain, and How to Prevent It
Stainless steel is stain resistant, not stain proof. Surgical instruments are made from hardenable martensitic stainless steel so that they can hold an edge and keep their spring. That steel resists corrosion very well, but only as long as its thin protective surface layer, called the passive layer, stays intact. Most of the spots and stains we see on instruments sent back to us were caused by something in the cleaning, rinsing or sterilizing process that damaged that layer. The good news is that once you know what each stain means, it is usually easy to fix the cause.
Reading the stain
Orange or rust-coloured spots
Often this is not the instrument rusting at all. It is rust or iron deposits carried over from somewhere else: an older, corroded instrument in the same tray, rust in the steam lines of the autoclave, or iron in the water supply. Wipe the spot with a pencil eraser or a soft cloth. If the mark comes off and the surface underneath is smooth, it was a deposit. If there is a small pit underneath, the steel itself has been attacked.
Pitting
Small pits that cannot be polished away are usually caused by chlorides. The most common sources are dried blood, saline, bleach and some disinfectants. Chlorides break through the passive layer at small points and eat into the steel. Pitting is permanent, and a pitted instrument should be retired, because the pits trap contamination.
Brown or black stains
These are often caused by detergents with the wrong pH, or by residue that was not rinsed off before sterilization and was then baked onto the surface in the autoclave.
Rainbow or blue-gold colours
A rainbow sheen usually means the instrument was overheated, for example in a dry heat cycle that was too hot, or that a detergent residue reacted with the surface. It is normally cosmetic but is a sign that the process needs checking.
White or light grey spots
These are mineral deposits from tap water used for rinsing or in the autoclave. They come off with a proper instrument cleaner and are prevented by rinsing with distilled or deionised water.
The main causes, and how to prevent them
- Dried blood and tissue. Keep instruments moist after use and clean them as soon as possible. Blood left to dry is the single most common cause of corrosion.
- Saline and bleach. Never soak instruments in saline or chlorine-based solutions, even briefly.
- The wrong detergent. Use a neutral or mildly alkaline enzymatic detergent made for surgical instruments, at the dilution the maker recommends.
- Tap water. Do the final rinse with distilled or deionised water, and make sure the autoclave uses the water quality its manufacturer specifies.
- Mixing metals. Do not process stainless steel together with damaged chrome-plated instruments or carbon steel items. Rust spreads from one to the other.
- Wet packs. Let the sterilizer complete its drying phase. Moisture trapped in a closed pack causes spotting during storage.
- Damaged instruments. Take corroded or pitted instruments out of service so they do not contaminate the rest of the tray.
What about new instruments?
New instruments are passivated during manufacturing, a chemical treatment that builds up the protective layer. The layer keeps renewing itself in normal use, as long as the instruments are cleaned and dried properly. Before first use, put new instruments through the full cleaning, lubrication and sterilization routine, the same as instruments returning from surgery.
Can stained instruments be saved?
Surface deposits and most stains can be removed with a stain remover made for surgical stainless steel, followed by thorough rinsing and lubrication. Abrasive pads, steel wool and household scouring powders should never be used, because they scratch the surface and make future corrosion more likely. Pitting, cracks at the box lock and jaws that no longer meet cannot be fixed by cleaning, and those instruments should be repaired or replaced.
For the full routine, see our step-by-step guide on how to clean and sterilize surgical instruments. Every instrument we make is produced from surgical-grade stainless steel and passivated before dispatch. Browse our sterilization and storage range, or ask us about a specific stain problem.